Cited 0 time in webofscience Cited 3 time in scopus

A 20V-compliance implantable neural stimulator IC with closed-loop power control, active charge balancing, and electrode impedance check

Title
A 20V-compliance implantable neural stimulator IC with closed-loop power control, active charge balancing, and electrode impedance check
Authors
Yao, L.[Yao, Lei]Zhao, J.[Zhao, Jianming]Li, P.[Li, Peng]Xue, R.-F.[Xue, Rui-Feng]Xu, Y.P.[Xu, Yong Ping]Je, M.[Je, Minkyu]
DGIST Authors
Je, M.[Je, Minkyu]
Issue Date
2015
Citation
2014 10th IEEE Asian Solid-State Circuits Conference, A-SSCC 2014, 201-204
Type
Conference
Article Type
Conference Paper
Abstract
An inductively powered implantable neural stimulator IC is presented in this paper. It features closed-loop power control, active charge balancing, and electrode impedance check functions. The stimulator IC is powered through 13.56MHz inductive link and supports 33.3kbps bi-directional telemetry with ASK for forward command transmission and LSK for backward data transmission, achieving 20V high compliance voltage, maximum 1.24mA stimulation current, and the resting potential of 50mV. The IC has an active area of 2mm×2mm implemented in 0.18-μm CMOS process with 24V LDMOS option. © 2014 IEEE.
URI
http://hdl.handle.net/20.500.11750/3680
DOI
10.1109/ASSCC.2014.7008895
Publisher
Institute of Electrical and Electronics Engineers Inc.