Detail View
A 0.05 mm3 diode-based single charged-particle real-time radiation detector for electron radiotherapy
WEB OF SCIENCE
SCOPUS
- Title
- A 0.05 mm3 diode-based single charged-particle real-time radiation detector for electron radiotherapy
- Issued Date
- 2025-04
- Citation
- Physics and Imaging in Radiation Oncology, v.34
- Type
- Article
- Author Keywords
- Single-particle sensitive dosimetry ; Electron EBRT ; Integrated circuit based dosimeter
- ISSN
- 2405-6316
- Abstract
-
Real-time radiation monitoring at the single-particle level is an unmet need for electron radiotherapy, especially for dose deposition to targets in motion or critical OARs. We have developed a first-in-class CMOS-based 0.05 mm3 single electron sensitive detector. The chiplet integrates all the requisite electronics. The functionality of the system is verified under 6 and 9 MeV clinical electron beams. Percentage depth vs. pulse-width curves for 6 and 9 MeV beams are measured and verified using Monte-Carlo simulations. The proposed system has the potential to enhance the electron radiotherapy quality and safety, providing real-time dosimetry from multiple sites simultaneously. © 2025 The Author(s)
더보기
- Publisher
- Elsevier
File Downloads
공유
Related Researcher
- Lee, Kyoungtae이경태
-
Department of Electrical Engineering and Computer Science
Total Views & Downloads
???jsp.display-item.statistics.view???: , ???jsp.display-item.statistics.download???: