Communities & Collections
Researchers & Labs
Titles
DGIST
LIBRARY
DGIST R&D
Detail View
Department of Robotics and Mechatronics Engineering
Bio-integrated Electronics Lab
1. Journal Articles
Thin Metallic Heat Sink for Interfacial Thermal Management in Biointegrated Optoelectronic Devices
Jung, Han Hee
;
Song, Juwon
;
Nie, Shuang
;
Jung, Han Na
;
Kim, Min Seok
;
Jeong, Jae-Woong
;
Song, Young Min
;
Song, Jizhou
;
Jang, Kyung-In
Department of Robotics and Mechatronics Engineering
Bio-integrated Electronics Lab
1. Journal Articles
Citations
WEB OF SCIENCE
Citations
SCOPUS
Metadata Downloads
XML
Excel
Title
Thin Metallic Heat Sink for Interfacial Thermal Management in Biointegrated Optoelectronic Devices
Issued Date
2018-11
Citation
Jung, Han Hee. (2018-11). Thin Metallic Heat Sink for Interfacial Thermal Management in Biointegrated Optoelectronic Devices. Advanced Materials Technologies, 3(11). doi: 10.1002/admt.201800159
Type
Article
Author Keywords
III-V optoelectronics
;
flexible electronics
;
heat dissipation
;
pulse oximetry
;
thermal damage
Keywords
STRETCHABLE ELECTRONICS
;
SKIN
;
SYSTEM
;
ARTIFACT
;
OXIMETRY
;
SILICON
;
DESIGN
;
SENSOR
;
BLOOD
;
ARRAY
ISSN
2365-709X
Abstract
The applications of modern optoelectronic devices have been extended, and they now provide practical means for seamless real-time monitoring of blood flow dynamics, by being integrated with flexible and stretchable wearable sensor platform technology. However, thermal management of these devices remains limited by undesired thermal energy originating from the heating of the light-emitting diode. Specifically, the surface temperature of the optoelectronic device becomes very high compared to that of the adjacent biological tissue, causing challenges in skin–optoelectronics integration and functional deterioration of the light source. In this study, an optoelectronic module that integrates the light-emitting diode, photodetector, and a thin metallic heat sink element for sustainable in situ thermal management is developed. Experimental and computational analysis results indicate that the proposed optoelectronic device has excellent heat dissipation capabilities for thermally safe long-term usability, due to the high thermal conductivity of the device and film-type geometrical design of the embedded heat sink for skin application. The proposed optoelectronic device architecture with metallic heat sink offers an ideal option for blood flow monitoring by providing both mechanical and thermal compatibility with biological tissue suitable for long-term clinical applications. © 2018 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
URI
http://hdl.handle.net/20.500.11750/9284
DOI
10.1002/admt.201800159
Publisher
Wiley-Blackwell
Show Full Item Record
File Downloads
There are no files associated with this item.
공유
공유하기
Related Researcher
Jang, Kyung-In
장경인
Department of Robotics and Mechatronics Engineering
read more
Total Views & Downloads